Brookfield YR-1 Rheometer User Manual

Page 9

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 9

Manual No. M02-215-B0412

United States

Outside United States

Hot (live)

Black

Brown

Neutral

White Blue

Ground (earth)

Green

Green/Yellow

I.3 Specifications

Speeds:

Pre Shear

0.01 to 200 rpm

Zero

0.01 to 0.5 rpm

Yield Test

0.01 to 5 rpm

Time Intervals

:

100 msec - 1000 msec

Weight:

Gross Weight

23 lbs. 10.5 kg.

Net Weight

20 lbs. 9 kg.

Carton Volume

1.65 cu. Ft. 0.05 m

3

Temperature sensing range:

-100°C to 300°C (-148°F to 572°F)

Analog Torque Output:

0 - 1 Volt DC = 0 - 100% Torque

Analog Temperature Output:

0 - 4 Volts DC = -100°C to +300°C

RS232 Compatible Serial Port for use with an attached printer or PC.

Centronics Compatible Parallel Port for use with an attached printer.

Torque Accuracy:

±1.0% of full scale range

Torque Repeatability:

±0.2%

Temperature Accuracy:

±1 °C : -100°C to +149°C

±2 °C : +150°C to +300°C

Operating Environment:

0 °C to 40 °C temperature range (32 °F to 104 °F)

20% - 80% R.H.: non-condensing atmosphere

Electrical Certifications:

Conforms to CE Standards:

BSEN 61326:

Electrical equipment for measurement, control and laboratory use - EMC

requirements

BSEN 61010-1:

Safety requirements for electrical equipment, for measurement, control

and laboratory use.