Figure III-18 is displayed when the yield test fails due to the resultant yield stress exceeding

the user defined limits. The limiting value causing the failed test (high or low yield stress) is

displayed in the first field to the right of the YLD: while the resultant yield stress measured

during the test is the second field.

LIMITS EXCEEDED

YLD:200 (726.36)

Figure III-18

After screen III-15 through III-18 is displayed, pressing the ESC key returns the screen to Fig-

ure III-8. At this point, after ensuring the appropriate spindle is attached, the Enter key can be

pressed to run the test again. Pressing the Enter key when any of these screens is displayed

immediately runs the test again and displays the first appropriate Run screen (Figures III-10

through III-13). When the test is complete, the results are sent out of both the parallel and se-

rial (RS-232) ports.

III.4 Data Output

III.4.1

Pre-Test Output

When a test is begun, the following information is sent out both the parallel and serial (RS-232)

ports in the format shown with each line terminated by a CR/LF (i.e., a Carriage Return/Line

Feed command):

Date

: ......................................................................

Sample Information

: ......................................................................

Operator

: ......................................................................

Model

: RV

YD1:Spindle

: 71/001

YD2:Pre-shear (rpm/secs)

: 0.1/0010

YD3:Zero (rpm)

: 0.05

YD4:Wait Time (secs)

: 0010

YD5:Run Speed (rpm)

: 0.4

YD6:Base Inc. (msec)

: 0200

YD7:Torque Red. (%)

: 020

YD8:Under-Range Red. (%)

: 050

YD9:Average (readings)

: 05

YDA:Limits (Low/High)(Pa)

: 00200/01000

Base Inc. Cal. (%)

: 1.55

Temperature (°C)

: 25.5

Test Name

: Sample Test Slot Number

: 01

Notes: 1) The Base Inc. Cal. (Base Increment Calibration) number is the change in torque

expected for the Base Inc. (Base Increment) selected using the calibration data

stored in the instrument.

2) The YDx values are for use with the EZ-Yield

TM

software and can be ignored.