Unpacking

Check carefully to see that all the components are received with no concealed damage.

(1)

Base, VS-2, with 2 Leveling Screws, VS-3, packed in a cardboard carton

(1)

Upright Rod, VS-34, with attached Clamp Assembly, VS-35Y

Assembly (Refer to Figure F-1)

1. Remove the base assembly from the carton.

2. Remove the screw and washer from the upright rod. Place the rod and clamp assembly into

the hole in the top of the base

.

Note: The “Front” designation on the clamp assembly should face the opening of the legs,

i.e., parallel to the leveling feet.

3. Rotate the rod/clamp assembly slightly until the slot on the bottom of the rod intersects the

pin located in the base.

4. While holding the rod and base together, insert the slotted screw and washer as shown and

tighten securely.

YR-1 Rheometer Mounting

Insert the YR-1 Rheometer mounting rod into the hole (with the cut-away slot) in the clamp

assembly. Adjust the instrument level until the bubble is centered from right to left and tighten

the clamp knob (clockwise). Use the leveling screws to “finely” adjust the viscometer level.

Note: If the Rheometer cannot be leveled, check to insure that the rod is installed with

the gear rack facing forward.

Note: If the clamp is taken off the upright rod, the tension insert (Part No. VS-29) must

be properly aligned for the clamp to fit back onto the upright rod. When the tension

insert (Part No. VS-29) is inserted, its slot must be in the vertical position parallel

to the upright rod. If the slot is not in the correct position, the clamp will not slide

down over the upright rod. Use a small screwdriver or pencil to move it into the

correct position. The VS-29W Belleville spring washers must face each other as

illustrated. Adjust the VS-28 tension screw so that the clamp assembly is not loose

on the upright rod.

Do not tighten the clamp knob unless the viscometer mounting rod is inserted

in the clamp assembly.

Center the YR-1 relative to the stand base and retighten the jam nut as required. Refer ring to

the Rheometer bubble level, adjust the leveling screws until the instrument is level.