Brookfield Engineering Labs., Inc.

Page 52

Manual No. M02-215-B0412

APPENDIX E - Communications

Communications

1. Serial (RS-232) Communications Parameters

Baud Rate

9600

Data Bits

8

Stop Bits

1

Parity

None

Handshake

None

2. RS-232 Output During The Test

All output that occurs immediately before and after the test is described in Section III.4. That

pre- and post- test output is sent out both the serial (RS-232) and parallel ports. When using

the Brookfield Computer Cable (Brookfield part # DVP-80), the YR-1 will output a data string

during the actual yield test run. This data is ONLY sent out to the RS-232 port for use with the

EZ-Yield

TM

software.

The following output is sent to the serial (RS-232) port at the end of every Base Increment period:

xxxxxx:yy.yy:ttt.t:zz.zz

where the delimiter between all fields is a colon (:) character and

xxxxxx = The Base Increment count in hexadecimal format, padded with leading zeroes.

This is the time, in milliseconds that each reading was taken in the instrument.

yy.yy

= % Torque value. % Torque is output with a resolution of 0.01% and a maximum

of 99.99%.

ttt.t

= Temperature in °C.

zz.zz

= Delta Torque (%). This is the actual change in percent torque, measured from the

previous reading, calculated at the end of each Base Increment period.

2

3

4

5

6

7

8

9

1

No Connection

Transmit Data (TxD)

Receive Data (RxD)

No Connection

No Connection

Serial Ground

Analog Ground

Analog % Torque

(Note 1)

Analog Temperature

(Note 2)

Notes:

1. This is a 0-1 volt d.c. output where 0 volts corresponds to 0% torque

and 1 volt corresponds to 100 % torque with a resolution of 1 millivolt

(0.1%).

2. This is a 0-4 volt d.c. output where 0 volts corresponds to -100 ºC and

4 volts corresponds to +300 ºC with a resolution of 1 millivolt (0.1 º C).

Figure E-1