If the digital display does not return to ZERO, the unit is in need of service (see Appendix I,

“How to Return Your Rheometer”).

Inaccurate Readings:

Verify spindle, speed and model selection.

Verify that %-Torque readings are greater than or equal to 10%.

Verify test parameters: Yield Test Parameters, container, volume and method.

Perform a calibration check; follow the instructions in Appendix B, “Calibration Pro-

cedures”.

Verify that tolerances are calculated correctly.

If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix I

for details on “How to Return Your Rheometer”.

Rheometer Will Not Return to Zero:

Rheometer is not level.

• Check with the spindle out of the sample.

• Adjust the laboratory stand.

Pivot point or jewel bearing faulty:

• Perform a calibration check. See Appendix C, “Calibration Procedures”.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair

Display Reading Will Not Stabilize:

Rheometer is not level

• Check with the spindle out of the sample.

• Adjust the laboratory stand.

Special characteristic of sample fluid. There is no problem with the rheometer.

Check for erratic spindle rotation.

• Verify the power supply.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair.

Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair.

Temperature fluctuation in sample fluid

• Use temperature bath control.

❏ No Recorder Response:

Verify the rheometer is not reading ZERO.

Verify the recorder is ON and not on STANDBY.

Verify the range settings.

Check cable leads for clean connection.