III.6 Test Results

III.6.1

Test Passed

When a yield test runs successfully, the resultant Yield Stress (Pa), the % Torque value at the

time of the yield and the Temperature at yield are printed.

Test (#5: Sample Tst) Complete

Yield Stress (Pa) = 726.36 % Torque @ Yield = 73.56 Temperature = 25.5C

In addition, the results are shown on the display as follows:

%=73.56

025.5C

YIELD-726.36

Pa

Figure III-21

III.6.2

Test Failed

Some of the reasons a yield test may fail are as follows.

III.6.2.1 Under-Range Condition

Only yield measurements in the % Torque range from 10% to 100% will be accepted. If

a Yield Stress is reached before the % Torque value is 10%, an Under-Range error causes

the test to end.

The following is sent to an attached printer:

Test (#5: Sample Tst) Complete

Test Failed = Under-range

The following screen is displayed:

YIELD ERROR
UNDER-RANGE

Figure III-22

III.6.2.2 Under-Range Error

The yield stress (Torque Reduction Value) is reached before the absolute % torque reaches

10%. When a test is started, once the absolute torque is greater than zero (0), the second,

third, and fourth Delta Torques (also known as the Base Increment Torque) are averaged.

The first three (3) Delta Torques after 9% absolute torque is reached are also averaged. An

under-range error occurs if the second average falls below the first average by 15% or more.