Maintenance and backup, Nucleoview™ software maintenance, Nucleocounter® instrument maintenance – ChemoMetec NC-3000 User Manual

Page 53

Advertising
background image

991-3003 Rev. 1.0

August 2013

53

Maintenance and Backup

NucleoView™ Software Maintenance

A two-channel image taken in an A8 slide with NucleoView NC-3000™ is approximately 15 MB in size.
Therefore, for example performing 50 viability and cell count assays each work day will over a month
accumulate approximately 16 GB.
The user must hence backup results and free up disc space in good time to ensure trouble free operation.
Each time a protocol run is started, NucleoView tests for sufficient disk space and if there is not enough, it
aborts the protocol with an error message to the user.
See

Analysis Results

for further details about file types and locations.


The NucleoView NC-3000™ constantly records all operations in event log files. These event log files should
be handled in the same manner as the image files with respect to backing up and freeing up disc space. The
event log files are always stored in the current results folder. One file is created for each month, and files
are named according to the rule <year><month>.log like 201205.log indicating a log file from May 2012.

The NucleoView NC-3000™ software including standard protocols is typically delivered on a USB stick, and
protocols and licenses are distributed as zip-files via e-mail, USB stick or other media.
The user must keep appropriate backup of these zip-files in order to be able to make a new installation in
case re-installation of the software is required.

NucleoCounter® Instrument Maintenance

ChemoMetec A/S provides specific Installation Qualification (IQ), Operation Qualification (OQ) and
Performance Qualification (PQ) protocols and associated test kits, which may be used to verify proper
operation of the NucleoView™ and NucleoCounter® NC-3000™ system.

The NucleoCounter® NC-3000™ instrument can be cleaned in case it has been contaminated with sample
fluids, dust or other spillage. Use dry cleaning swabs for cleaning, and only if this is not enough, use 96-
100% pure ethanol.
The outer surfaces of the NucleoCounter® NC-3000™ instrument may also be cleaned with 96-100% pure
ethanol if needed.

It is good practice to insert the sample tray, whenever the instrument is not being used, to prevent dust
build up on the internal surfaces.

Advertising