Trouble Shooting

Installation
The NucleoView NC-3000™
software or the NucleoCounter®
NC-3000™ instrument is not
installed correctly.


The user must be logged on with administrator rights during
installation.
When connecting the NucleoCounter® NC-3000™ instrument to a
previously unused USB port, the NucleoView NC-3000™ software may
need to reinstall the camera drivers and will prompt the user to do so.
This operation may need administrator rights, so either use another
USB port where the camera has already been installed, or log on with
administrator rights and complete the camera installation as
described in Installation of Instrument.

NucleoView NC-3000™ Issues
(Software)

Protocol file is invalid



Connection error



The license file is corrupted, missing or does not include a license for
the selected protocol. The appropriate license must be installed as
described in the Licenses section.

The NucleoCounter® NC-3000™ instrument icon in the upper left
corner of the main window stays red after NucleoView NC-3000™ has
been launched, i.e. the NucleoView NC-3000™ software does not
connect to the instrument. This may be caused by an error in the USB
connection. To inspect and solve this, open the Control Panel and
launch System. Select the Hardware tab and click Device Manger. In
case of a USB driver error there will be a USB item marked with a
yellow dot containing an exclamation mark. Select properties for this
item and select the driver tab. Select Update Driver and let the system
select a driver automatically.

Cassette

No cassette is found


Non-valid cassette



Flow detection error


There is no cassette in the sample tray.
Place a cassette which has been correctly loaded into the sample tray.

A used cassette has been placed in the sample tray.
Remove the used cassette and replace with a new correctly loaded
cassette.

The flow has failed to reach the flow detection point within given time
limits.
Too little sample volume or bubbles have been loaded into the
cassette. Replace cassette with a new, correctly loaded cassette.