The Compex Wireless stimulator needs special precautions regarding EMC and needs to be installed and put into service according to the EMC

information provided.

Portable and mobile RF communications equipment can affect the Compex Wireless stimulator.

The use of Accessories, transducers, and cables other than those specified by the manufacturer, may result in increased Emissions or decreased

Immunity of the Compex Wireless stimulator.

The Wireless stimulator should not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary, the Compex

Wireless stimulator should be observed to verify normal operation in the configuration in which it will be used.

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS

The Compex Wireless stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of the

Compex Wireless stimulator should assure that it is used in such an environment.

EMISSION TESTS

COMPLIANCE

ELECTROMAGNETIC ENVIRONMENT – GUIDANCE

Radiated Emissions CISPR 11

Group 1

The Compex Wireless uses RF energy only for its internal function.

Therefore, its RF emissions are very low and are not likely to cause

any interference in nearby electronic equipment.

Conducted Emissions CISPR 11

Class B

The Compex Wireless is suitable for use in all establishments,

including domestic establishments and those directly connected to

the public low-voltage power supply network that supplies buildings

used for domestic purposes.

Harmonic emissions IEC 61000-

3-2

Class A

Voltage fluctuations IEC 61000-

3-3

Complies

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS

The Wireless stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of the Wireless

stimulator should assure that it is used in such an environment.

IMMUNITY TEST

IEC 60601

TEST LEVEL

COMPLIANCE LEVEL

ELECTROMAGNETIC

ENVIRONMENT - GUIDANCE

Electrostatic discharge (ESD)

IEC 61000-4-2

±6kV contact

±8kV air

±6kV contact

±8kV air

Floors should be wood, concrete

or ceramic tile. If floors are

covered with synthetic material,

the relative humidity should be

at least 30%.

Electrical fast transient/burst

IEC 61000-4-4

±2kV for power

supply lines

±1kV for

input/output lines

±2kV (power lines)

Not Applicable (I/O lines)

Mains power quality should be

that of a typical commercial or

hospital environment.

Surge

IEC 61000-4-5

±1kV Line to Line

(or Neutral)

±2kV line(s) to earth

±1kV Line to Line

Not Applicable

(Line to Earth)

Mains power quality should be

that of a typical commercial or

hospital environment.

Voltage dips, short interruptions

and voltage variations on power

supply input lines

IEC 61000-4-11 a

<5% UT (>95% dip in UT) for 0,5

cycle

40% UT (60% dip in UT) for 5

cycles

70% UT (30% dip in UT) for 25

cycles

<5% UT (>95% dip in UT) for 5 sec

<5% UT (>95% dip in UT) for 0,5

cycle

40% UT (60% dip in UT) for 5

cycles

70% UT (30% dip in UT) for 25

cycles

<5% UT (>95% dip in UT) for 5 sec

Mains power quality should be

that of a typical commercial

or hospital environment. If the

user of the Compex Wireless

stimulator requires continued

operation during power mains

interruptions, it is recommended

that the Compex Wireless

stimulator be powered from an

uninterrupted power supply or

a battery.

Power frequency (50/60Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields

should be at levels characteristic

of a typical location in a

typical commercial or hospital

environment.

a

U

T

is the a.c mains voltage prior to application of the test level.