6 TROUBLESHOOTING

ELECTRODE FAULT

The remote control shows the

symbol of an electrode and a

disconnected module on the

channel where the problem

is detected (in this example,

channel 1).

Check whether the electrodes

are properly connected to the

module.

Check whether the electrodes

are old, worn, and/or the

contact is bad: If you suspect

this may be the cause, try using

new electrodes.

MODULE OUT OF RANGE

The remote control shows the out of

range symbol on the channel where

the problem has been detected (in

this example, channel 1).

Check to see whether the

module and the remote are less

than 2 metres apart.

Make sure that you are not

in a remote location with any

obstacles which can reflect the

signal of the remote control.

Make sure that you are in a

place where the signal of the

remote control can be reflected.

SYNCHRONISATION PROBLEM

If the synchronisation process

has been interrupted or if for any

reason you cannot synchronise

(remote disconnected, power

failure, etc.), the remote control

may, display this screen.

Reconnect the remote control

to your PC and re-start the

synchronisation process by

pushing the button for channel

4 (the one with the

icon).

THE MODULE’S BATTERY HAS RUN OUT

During stimulation, a module’s

battery may run out. If this

happens, the symbol of an

empty battery is displayed on

the channel where the problem

has been detected (in this

example, channel 4).

You can stop the stimulation

and recharge the module with

the empty battery.

You can set aside the module

with the empty battery and

continue your stimulation

session without it.

MODULE LED BEHAVIOUR

The LED blinks green and red alternately: the module is out of range or not recognised by the remote control.

Make sure the remote control is switched on.

Check to see whether the module and the remote are less than 2 metres apart.

THE LED IS ALWAYS RED

Make sure the module is charged.

Try to re-start the remote control and the modules.

If despite this the LED is still red, contact an official Compex customer service centre.

STIMULATION IS NOT PRODUCING THE USUAL SENSATION

Check that all the settings are correct and ensure the electrodes are positioned properly.

Change the positioning of the electrodes slightly.

THE STIMULATION CAUSES DISCOMFORT

The electrodes are losing their bonding strength and are no longer providing suitable contact with the skin.

The electrodes are worn and need to be replaced.

Change the positioning of the electrodes slightly.

THE UNIT IS NOT WORKING

Make sure the remote control and the modules are switched on.

Try to re-start the remote control and the modules.

If despite this the unit still does not work, contact an official Compex service centre.