6. MEASUREMENT

6-1 Preparing the Sample (For the SV-A series)

* Vibration or shock during transportation may cause a display value error. Before measurement,

calibrate the viscometer using the sample cup that will be used for measurement. For calibration
with purified water, refer to “8-2-3 Simplified Calibration Using Purified Water”. (Only for SV-1A/10A)

Model: SV-1A

1 Pour the sample into the cup until its surface is between the level

gauges.

- With the 2 mL sample cup, the lower level gauge indicates 1.8

mL, the upper level gauge indicates 2 mL.

- With the 45 mL sample cup, the lower level gauge indicates 35

mL, the upper level gauge indicates 45 mL.

*The glass sample cup does not have level gauges.

2 Attach the cup on the table along the guides.

- With the 2 mL sample cup, place the sample cup in the sample

cup holder (For 2 mL), and set it into the 45 mL sample cup.

* When using the sample cup holder, select either the transparent

holder or the black holder, depending on the sample color.

3 Confirm that the protector is in the position as shown in the

figure.

Raise the lever to release the sensor unit.

4 Pinch the grips, support the front side of the sensor unit and

gently lower the sensor plates above the sample surface.

Adjust the X-Y-Z stage if necessary.

5 Lower the lever to secure the sensor unit.

6 Confirm that the positioning of the sensor plates is in center of

the sample cup by adjusting the X-Y-Z stage.

- When using the 2 mL sample cup, raise the protector.

Sample cup
(2 mL)

Sensor

plate

Protector

Level

gauges

Lever

Grips

Sensor unit

Guides

Sample cup

<2 mL>

Sample cup

holder

<For 2 mL>

Sample cup

<45 mL>