A&D Weighing SV-100H User Manual

Page 49

Advertising
background image

47

9 When the measurement of the first point has

completed, clean the sensor plates,
temperature sensor and protector and
prepare the second standard viscosity fluid.

10 Place the second standard viscosity fluid in

the sample cup. Press the START key to
start the measurement of the second point.

11 After the measurement, wait for the display

to become stable and press the PRINT key.
The measurement value blinks and is ready
to be corrected.

12 Correct the value using the following keys:

MODE key

Switches the blinking digits.

START key

Changes the setting of a
blinking digit.

STOP key

Moves the decimal point.

13 Press the PRINT key to confirm the

correction value.

14 To exit the calibration mode:

With " end " being displayed, press the PRINT
key again. The calibration data is saved and the
display returns to the standby mode.

To correct the value:
With " end " being displayed, press the

STOP key and correct the value.


15 Measure the viscosity of the two standard

viscosity fluids used. Confirm that the
viscometer displays the similar value as the
entered correction value for each fluid.

From the previous page

Set standard viscosity fluid

(Second point measurement mode)

When the data

is stabilized

(Correct the second point value)

(End of two-point calibration)

To correct

(Standby mode)

(Confirm the value

of the first fluid)

(Confirm the value

of the second fluid)