91

AX-SV-37 Water Jacket Assembly

• Used, in combination with a commercially available

constant temperature bath for heating medium
circulation, to maintain the sample temperature
constant or to measure the viscosity while changing
the sample temperature.

• The glass sample cup sold separately can also be

used.

• The AX-SV-37 consists of :

Water jacket

1 pc

(Main body: Polycarbonate, Packing: Silicone rubber,
Washer: Nylon)

Small sample cup

4 pcs

(Polycarbonate, Operating temperature: 120°C max.)

Small sample cup cover

4 pcs

(Polycarbonate, Operating temperature: 120°C max.)

Screw 1

pc

(Screw: Polyacetal, Washer: Nylon)

• Specifications Circulation nozzle: Outside diameter 10.5 mm

Recommended hose: Silicone tube, inside diameter 8 mm

• It is recommended that, for safety, a commercially available tubing clamp (clamping size: 11 to

20 mm) be used to fasten the tube securely to the nozzle.

• When using the water jacket, make sure that no inner pressure is exerted in the water jacket

due to the kinked or blocked tubes, as that could exert the pressure in the water jacket,
causing it to break.

• When a stirrer is attached at the bottom of the water jacket, the sample viscosity can be

measured while the sample is being stirred. The maximum viscosity value which can be
measured is 1000 mPa

⋅s.

Stirrer: VARIOMAG MICRO manufactured by H+P Labortechnik AG

Use a rotator with a size of 6 mm (length) x 4 mm (diameter).