Example of the function setting procedure

To change the setting of "Unit upon power-on (Unit)" to the viscosity: cP (Centipoise) and the

temperature: °C (Celsius).

1 In the standby mode, press and hold the MODE

key to enter the function setting mode. " 5vtyp "
appears.

2 Press the MODE key to select " Unit ".

3 Press the PRINT key to confirm the item.

(The decimal point illuminates when the setting
currently saved is displayed.)

4 Press the START key or HOLD key to select the

unit to be used.

(In this example, " 2 " is selected. Viscosity: cP,
Temperature: °C)

5 Press the PRINT key to save the setting.

After " end ", the next item is displayed.

6 Press the STOP key to return to the standby

mode.


(Standby mode)

Press and hold

or

Next item of

function setting

(Standby mode)