A&D Weighing SV-100H User Manual

Page 32

Advertising
background image

30

7 Turn the knob (Height direction) so as to

adjust the sample surface to the center of
the narrow part of the sensor plates.
Confirm that the sample surface is between
the upper and lower triangular marks.

Note

When using the 2 mL sample cup, raise

the protector.

Adjust the height of the sample surface

correctly, otherwise it may cause a
measurement error.


Model: SV-10A / SV-100A

1 Pour the sample into the cup until its surface is between the

level gauges.

- With the 45 mL sample cup, the lower level gauge indicates

35 mL, the upper level gauge indicates 45 mL.

- The 13 mL glass sample cup does not have level gauges.

The position approximately 25 mm above the cup bottom
indicates 13 mL.

2 Attach the cup on the table along the guides.

3 Confirm that the protector is in the position as shown in the

figure.

Raise the lever to release the sensor unit.

4 Pinch the grips, support the front side of the sensor unit and

gently lower the sensor plates above the sample surface.

5 Lower the lever to secure the sensor unit.

Sample cup (45 mL)

Guides

Protector

Sensor

plate

Knob

(Height direction)

Sensor protective

cover

Protector

Sample

surface

adjustment

position

Glass sample cup (13 mL)

Level gauges

Approx. 25cm

Advertising