8-2-3 Simplified Calibration Using Purified Water (SV-1A/1H/10A/10H)

1 Place the purified water in the sample cup.

2 Press the START key to measure

the purified water. Confirm that the
viscosity and temperature values are
stabilized.

3 Press and hold the START key.

The theoretical viscosity value

(Viscosity

× Density)

of the purified

water at the measuring temperature
is displayed and all the displays
blink.

To cancel the operation, press the

STOP key. The display returns to

the status before calibration.

4 Press the START key again to

perform calibration.
When calibration is completed, " end "
is displayed. Then, the viscosity
returns to the measuring status.