Atomic calibration commands, Clv command, Cli command – Maxim Integrated 6613_OMU_2+2S_URT_V100 User Manual

Page 14: Clp command

Advertising
background image

6613_OMU_2+2S_URT_V1_00 Firmware Description Document

UG_6613_060

14

Rev. 1.1

4.5.2 Atomic Calibration Commands

The atomic calibration commands provide individual calibration of:

• Voltage.
• Current.
• Phase.
• Temperature.
• Power.

A sequence of these commands results in full calibration of the unit.

4.5.2.1 CLV Command

The CLV atomic calibration command calibrates voltage to the target value and tolerance and saves the
coefficients to flash. To calibrate the voltage, enter the CLV command:

>CLV<CR>

The response is:

VCal OK:
>

4.5.2.2 CLI Command

The user can then calibrate the current using the CLI command. The CLI command calibrates the current
from the specified channel to the target value and tolerance and saves the coefficients CLW. To calibrate
the current for channel 1, enter the following:

>CLI<CR> or CLI1<CR>

The response is:

ICal 1 OK:
>

The CLI2 command performs the current calibration for channel 2.

The CLI3 command performs the current calibration for both channel 1 and channel 2.

4.5.2.3 CLP Command

The user can calibrate for phase added by a current transformer by using the CLP command. The CLP1
command calibrates the phase from channel 1 to the target value and tolerance and saves the coefficient
to flash. An example of the procedure is given below.

Apply a controlled precision voltage and current signal at a set phase angle.

1. Enter target phase angle at )C3.
2. Enter phase tolerance at )BF
3. Enter phase calibration command.

>CLP<CR> or CLP1<CR>

The response is

>PCal 1 OK:

Advertising