UG_6613_ 060

6613_OMU_2+2S_URT_V1_00 Firmware Description Document

Rev. 1.1

31

Reserved

B0 - BC

0

Reserved

Reserved

Additional

Status

BD

1

Bit 0 – Reserved.

Bit 1 – WPULSE Disable.

Bit 2 – VCal Failure.

Bit 3 – ICal1 Failure.

Bit 4 – WCal1 Failure.

Unused

BE

Tolerance on

Phase

BF

0.001°

0.100°

Measured value to fall within this

set tolerance of the target value

(Calibration Current entry) for the

calibration to be complete.

If the tolerance to the target

phase is desired to be more

coarse, to within 0.5°, the

user can enter the

following:

>)BF=+0.500<CR>

Reserved

C0

0

Reserved

Reserved

Calibration

Voltage

C1

mVrms

+120.000

Target line voltage (rms) used for

calibration.

If the target line voltage for

calibration is 220V, enter

the following:

>)C1=+220<CR>

MPU

Parameter

Location

(hex)

LSB

Default

Comment

Example

Calibration

Current

C2

mArms

+1.000

Target load current (rms) used for

calibration.

If the target load current for

calibration is 2A, enter the

following:

>)C2=+2<CR>

Calibration

Phase

C3

0.1°

+0

Target Phase (voltage to current).

Normally set to zero.

Tolerance on

Voltage

C4

mVrms

+0.010

Measured value to fall within this

set tolerance of the target value

(Calibration Voltage entry) for the

calibration to be complete.

If the tolerance to the target

voltage is desired to be

more coarse, to within

0.1V, the user can enter the

following:

>)C4=+0.100<CR>

Tolerance on

Current

C5

mArms

+0.010

Measured value to fall within this

set tolerance of the target value

(Calibration Current entry) for the

calibration to be complete.

If the tolerance to the target

current is desired to be

more coarse, to within

0.1A, the user can enter the

following:

>)C5=+0.100<CR>

Average

Count for

Voltage

C6

1

+3

Number of voltage measurements

taken and averaged to be

compared to the target value

(Calibration Voltage entry).

If the amount of averaging

for the voltage

measurement is desired to

increase to 10 enter the

following:

>)C6=+10<CR>

Average

Count for

Current

C7

1

+3

Number of current measurements

taken and averaged to be

compared to the target value

(Calibration Current entry).

If the amount of averaging

for the current

measurement is desired to

increase to 10 enter the

following:

>)C7=+10<CR>