Operation bvc professional / bvc professional g – VACUUBRAND BVC professional G User Manual

Page 24

Advertising
background image

page 24 of 68

Operation BVC professional / BVC professional G

Do not press the keys. Only touch the keys of the touch panel.

The keys “+“ and “-“ have to be touched > 0.25 seconds to be actuated.

The other keys have to be touched > 0.5 seconds.

Touch the keys with LED

below the LED.

A successful action is confirmed by a blip and the flashing of LEDs.

Use the keys to set suction power of the system.

The suction power can be set as underpressure (relative to atmo-

spheric pressure) linearly in a range of 150 mbar (1 LED flashes)

to 850 mbar (8 LEDs flash).

+

-

From higher software versions

A

flashing LED indicates that the actual avail-

able suction power differs from the preset suc-

tion power.

Attention: If the collection bottle is under vacu-

um and the vacuum demand is reduced, the ex-

isting underpressure inside the bottle remains

until the vacuum is reduced (pressure rises) by

further aspiration with the VHC

pro

.

Position

Designation

1

Key to select bottle size and

level sensor

2

Key to reduce suction power

3

Key “bottle change”

4

Display suction power

5

Key to increase suction

power

4
5

2

1

3