Troubleshooting, Fault possible cause remedy – VACUUBRAND BVC professional G User Manual

Page 50

Advertising
background image

page 50 of 68

Fault

Possible cause

Remedy

❑ Pump fails to start,

system becomes

vented.

➨ Not plugged in, or failure

of electrical supply?

✔ Plug in, or check fuse

controlling power outlet.

➨ Power switched off?

✔ Switch on power.

➨ Motor overloaded?

✔ Switch off, allow motor

to cool, identify cause of

failure and eliminate be-

fore switching on again.

➨ Internal fuse blown?

✔ Identify cause of failure,

replace fuse (under the

cover).

➨ Aspiration power setting

is off (atmosphere, no

vacuum)?

✔ Set aspiration power to

“+“ (more vacuum).

➨ Red LED glows continu-

ously?

✔ Eliminate leaks.

➨ Filter clogged?

✔ Replace filter.

❑ No aspiration at the

VHC

pro

, pipette drips

(above causes ex-

cluded).

➨ Adapter or aspiration

tube defective?

✔ Replace adapter with

aspiration tube.

➨ Filter clogged?

✔ Replace filter.

➨ Leak at the bottle head? ✔ Check seals, coupling

and blind cap. Screw in

or replace if necessary.

➨ Pump operating but no

pumping speed?

✔ Perform maintenance,

replace diaphragm and

valves if necessary.

➨ Aspiration power setting

is off (atmosphere, no

vacuum)?

✔ Set aspiration power to

“+“ (more vacuum).

❑ Pump switches

frequently.

➨ Leak in the system?

✔ Check tubing, seals,

coupling, and blind cap.

Screw in or replace if

necessary.

➨ Filter clogged?

✔ Replace filter.

➨ Diaphragms or valves of

the pump defective?

✔ Perform maintenance,

replace diaphragm and

valves if necessary.

Troubleshooting