Disinfection routine, Notice – VACUUBRAND BVC professional G User Manual

Page 33

Advertising
background image

page 33 of 68

Emptying the collection bottle of the BVC professional

and BVC professional G

The

level sensor switches off the pump at a

liquid level of approx. 80% of capacity if cor-

rectly adjusted and if the correct bottle size is

selected. This is indicated by “blips” and by a

red LED inside of the bottle symbol on the “bot-

tle change” key. Due to the underpressure in-

side the bottle, further aspiration is possible for

a short time.

Touching the key below the LED switches off

the blips and activate the pump, the LED is still

flashing.

It is possible to aspirate for 3 minutes more, un-

til the alarm is activated again and the pump is

switched off again, e. g. for suction of disinfec-

tants. Attention: Ensure to not overfill the bottle.

The action is repeatable.

Check liquid level in the collection bottle regu-

larly.

Maximally admissible liquid level in collec-

tion bottle: approx. 80 %, depending on the

application. (In case of low boiling liquids or in

case of suction of liquids which tend to foam,

the admissible liquid level might be reduced.)

Disinfection routine

or

Touch the key prior to changing the collection

bottle (red LED changes from flashing to glow-

ing) or switch off the BVC, to avoid running of

the pump while no bottle is connected.

When removing the tubing at the pump, the

pump stops automatically (only BVC profes-

sional).