Ensure that the equipment and all components are suit-

able for the intended application. Use the system only for

aspiration of liquids and filtration.

Setting up and installing the system

Equipment must be connected only to a

suitable elec-

trical supply and a suitable ground point. Failure to

connect the motor to ground may result in deadly elec-

trical shock.

The supply cable may be fitted with a molded Euro-

pean IEC plug or a plug suitable for your local electri-

cal supply. If the plug has been removed or has to be

removed, the cable will contain wires color coded as

follows: green or green and yellow: earth; blue or white:

neutral; brown or black: live. The device has an internal

fuse.

+

Do not permit any

uncontrolled pressurizing (e.g., make

sure that an exhaust tubing cannot become blocked).

Risk of bursting!

+

Due to the high compression ratio of the pumps, pres-

sure at the outlet port may be generated that is higher

than the maximum permissible pressure compatible

with the mechanical stability of the system.

+

Always provide a free and pressureless exhaust line.

• Comply with

maximum permissible pressures and

pressure differences; see section ”Technical data”. Do

not operate the pumping unit with overpressure at the

inlet or outlet.

• Check that line voltage and current conform with the

equipment. (See rating plate.)

• Avoid overpressure of more than 0.2 bar in instances in

which inert gas is connected.