Fault

Possible cause

Remedy

❑ Pump is running

continuously.

➨ Leak in the system?

✔ Check tubing, seals,

coupling, and blind cap.

Screw in or replace if

necessary.

➨ Too high underpressure

for altitude / weather

conditions?

Vapor pressure of the

media inside the collec-

tion bottle too high?

✔ Reduce underpressure /

aspiration power.

➨ Other causes?

✔ Contact your local

distributor.

❑ Keys do not react.

➨ Keys of the touch panel

touched long-lasting?

✔ Wait 20 seconds and try

again.

❑ Only BVC profes-

sional / G: Level

sensor gives a false

alarm.

➨ Bottle has been re-

placed or changed am-

bient conditions?

✔ Adjust the level sensor.

➨ Not using original

VACUUBRAND BVC

bottle?

✔ Use original

VACUUBRAND BVC

bottle or original spare

bottle.

➨ Adhesive foil at the col-

lection bottle or the level

sensor foil?

✔ Remove adhesive foil.

➨ Bottle wet (after clean-

ing or disinfection)?

✔ Dry bottle. Adjust the

level sensor.

➨ Bottle inside and/or

outside with conductive

coating (e.g., silver con-

taining disinfectants)?

✔ Clean bottle carefully

and dry. Adjust the level

sensor.