Adjustment routine for BVC professional with 4l polypropylene bot-

tle from software version V1.6, see rating plate

Action

Reaction

Meaning

Step 1

Touch and hold “4l P” key.

Case 1:

The LED of the “4l PP“

bottle symbol flashes

blue.

Adjustment is possible by

continuing to hold the “4l

PP” key, and proceeding

to step 2.

Case 2 / false alarm:

The alert tone for the false

alarm is shut off.

The LED of the “4l PP“

bottle symbol flashes

blue.

Step 2

When the LED of the “4l

PP” key flashes blue,

touch and hold the “bottle

change” key at the same

time.

A one-time buzzer

sounds.

After the buzzer, the LED

of the “bottle change” key

lights red.

Adjustment has been car-

ried out correctly.

Step 3

Release both “4l PP” key

and “bottle change” key. The LED of the “4l PP”

key lights blue.

The red LED of the “bottle

change” key turns off.

Adjustment finished.

Attention: Do not interrupt the adjustment routine once you have begun,

even if there is an alarm. Complete the entire routine, and repeat if neces-

sary.