With a new bottle, or to correct false alarms – VACUUBRAND BVC professional G User Manual

Page 38

Advertising
background image

page 38 of 68

At the factory, the level sensor is adjusted either to the 4l polypropylene

bottle or to the coated 2l glass bottle (G version) at standard conditions.

It may be necessary to adjust the sensor to compensate for changes in

bottle material, bottle size or changed ambient conditions.

Adjustment of the BVC level sensor to

work properly with a new bottle, or to

correct false alarms

NOTICE

Preparation

NOTICE

NOTICE

Use only the original VACUUBRAND BVC bottle or the original spare bot-

tle (see “Accessories - Spare parts”).

1. Do not switch off the BVC.

2. To shut off a false alarm prior to an adjustment, simply touch the

key marked “bottle change.” The red LED will continue to flash.

3. Empty the bottle.

4. Set the correct bottle size at the touch panel.

5. Wait approx. 1 minute before adjustment.

We also recommend emptying the bottle and adjusting the BVC profes-

sional if there is a false alarm indicating “liquid level” even though the

bottle is less than 3/4 full.

If there is a false alarm in spite of an adjustment, repeat the adjust-

ment.

If there is a false alarm in spite of

repeated adjustment, please call

+49 9342 808-5500 (after sales service), or your local VACUBRAND

sales office.

Pertains only to the BVC professional / BVC professional G