Sekusept

®

Plus

The use of the disinfectant

Sekusept

®

Plus

(Manufacturer: Ecolab GmbH & Co OHG, Düs-

seldorf, Germany) in extensive in-house test-

ing did not cause any damage to the collection

bottle. Hence the disinfectant Sekusept

®

Plus

- even when used in the collection bottle dur-

ing suction - is preferred to chlorine bleach.

Comply with use and safety instructions of the

manufacturer!

+

Even if using disinfecting solutions

after termination of the suction it

is absolutely necessary to assure the compatibility of disinfectant and

the parts to be disinfected.

+

For information about the compatibility with the materials of the pump-

ing unit, ask the manufacturer of the disinfectant.

+

The wetted materials of the system are listed in the section ”Technical

data”.