Safety during operation – VACUUBRAND PC 8 m. RC 6 User Manual

Page 17

Advertising
background image

page 17 of 49

reactions of the pumped substances with each other

and with the pump’s materials.

Safety during operation

Adopt suitable measures to prevent the release of dan-

gerous, toxic, explosive, corrosive, noxious or polluting

fluids, vapors and gases. To prevent any emission of

such substances from the pump outlet, install an ap-

propriate collecting and disposal system and take pro-

tective action for pump and environment.

You must take suitable precautions to prevent any for-

mation of explosive mixtures in the oil reservoir (PC 8

with RC 6), in the pump chamber, or at the outlet of the

pump. In case, e.g., of a diaphragm failure, mechani-

cally generated sparks, hot surfaces or static electricity

may ignite these mixtures. Use inert gas for gas ballast

or venting, if necessary.

Drain appropriately or otherwise remove any potential-

ly explosive mixtures at the outlet of the pump, or dilute

them with inert gas to non-explosive concentrations.

Never operate this pump if it has a damaged cord or

plug.

+

If the pump is not working properly, has been dropped

or has fallen into water, contact your pump service pro-

vider.

+

Prevent any part of the human body from coming into

contact with vacuum.

+

Never aspirate liquids or dust into the pump.

+

Make sure that the exhaust pipeline cannot become

blocked.

+

Check the overpressure safety relief device at the ex-

haust waste vapor condenser at appropriate intervals.

Advertising