inert gas supply is connected to the gas ballast, ensure

that its inlet pipeline is not contaminated.

• Pay attention to the safety symbol ”hot surfaces” on

the equipment. Hot parts may cause burns if touched.

Adopt suitable measures to prevent any danger arising

from hot surfaces or electric sparks. Ensure that hot

surfaces of the pump do not cause burns. Provide a

suitable contact guard if necessary.

• Ensure that the

coolant outlet pipe at the waste vapor

condenser is always free and that it cannot get blocked.

Do not start the pump if the pressure difference between

inlet and outlet exceeds 16 psi (1.1 bar) at maximum.

Prevent the backpressure of gases and the backflow of

condensates at the outlet.

Check the liquid level in the collecting flask regularly and

drain condensate in time to prevent overfilling.

Provide appropriate protective measures to allow for the

possibility of failure and

malfunction. The protective mea-

sures must also allow for the requirements of the respec-

tive application.

In case of overload, the motor is shut down by a

self-hold

thermal cutout in the winding.

Note: Only manual reset is possible. Switch off the pump

and disconnect from the power source. Identify and elimi-

nate the cause of failure. Allow the pump to cool down suf-

ficiently before restart. Wait approx. five minutes before

restarting the pump.

Maintenance and repair

In order to comply with laws (occupational, health and

safety regulations, safety at work law and regulations for

environmental protection) vacuum pumps, components

NOTICE

NOTICE