Installation of upgrade kit pc 8, Notice – VACUUBRAND PC 8 m. RC 6 User Manual

Page 37

Advertising
background image

page 37 of 49

The upgrade kit PC 8 allows the mounting of either the Chemistry-

HYBRID pump RC 6 or the chemistry diaphragm pump MD 12C / MV

10C.

+

Consequently, the upgrade kit contains parts that are

not needed in

individual cases.

Attention: If the HYBRID pump RC 6 contains oil, take

great care not to tilt the pump to an angle that would allow

oil to make its way into the filter element.

+

Lay the pump on its side: if installing with diaphragm pump MD 12C /

MV 10C, the rating plate of the pump has to be on top, if installing the

HYBRID pump RC 6, the oil mist filter must be on top (support pump

appropriately).

Note that to install the console, the exhaust waste vapor condenser

and the collecting flask must be on top.

+

Remove the shock absorbers from the pump’s feet and install the shock

absorbers which have male threads at both ends. The original shock

absorbers are secured with adhesive to prevent them from working

loose.

+

Push the console over the threaded shock absorbers. Note that there

are two sets of bores for the two different pump types.

NOTICE

bores for MD 12C / MV 10C

bores for RC 6

console

Installation of upgrade kit PC 8

Advertising
This manual is related to the following products: