Exhaust waste vapor condenser:

Assemble the hose nozzles for coolant inlet

(13) and coolant outlet (14) tubing at the ex-

haust waste vapor condenser.

Assemble the collecting flask at the exhaust

waste vapor condenser using a joint clip.

The

exhaust waste vapor condenser enables an effi-

cient condensation of the pumped vapors at the outlet.
+

No backflow of condensates.

+

Controlled recovery of condensates.

+

Close to 100% solvent recovery.

+

The isolation cover protects against glass splinters in

case of breakage, acts as thermal isolation to avoid

condensation of humidity and is intended to absorb

shocks.

Attach the tubing of the coolant circuit to the respective

hose nozzles (hose nozzles for tubing I.D. 1/4”-5/16”

(6-8 mm), see image) at the waste vapor condenser.

Check all hose connections prior to starting operation of

the cooling system.

Secure coolant hoses at the hose nozzles (e.g., with hose

clamps) to prevent their accidentally slipping off.

Prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere. If appropriate, connect

the exhaust line to a suitable treatment system.

+

Never block the gas outlet (hose nozzle for tubing I.D.

3/8” (10 mm)). The exhaust hose has always to be un-

obstructed and without back pressure to enable an un-

hindered discharge of gases and protect the pump from

damage.

Note: Install the hoses of the cooling system in a way

to avoid the flow / dripping of condensed water onto the

pumping unit (especially cables and electronic parts).

13

14

NOTICE