• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

Keep a distance of minimum 8 in (20 cm) between fan and

adjacent equipment or casework.

Use connecting hoses with large diameter and keep them

as short as possible to avoid flow losses. Locate the pump

as closely as possible to the application.
Always install outlet tubing descending from the pump to

avoid backflow of condensate towards the pump.

Use a suitable valve to isolate the pump from the vacuum

application. This is to allow the pump to warm up before

pumping condensable vapors and to clean the pump after

use before it is switched off.

When assembling, ensure

vacuum-tightness. After as-

sembly, check the whole system for leaks.

Secure hose connections at the pump appropriately, e.g.,

with hose clamps, to protect against accidental detach-

ment.

To reduce pump noise emanating from the pump exhaust

port, connect an exhaust hose or use a silencer.

Separator (inlet) and

exhaust waste vapor condenser (outlet)

Separator at inlet (accessory; only with dia-

phragm pump)

:

The separator at the inlet protects against drop-

lets and particles from entering the pump.
+

Enhances lifetimes of diaphragms and

valves.

+

Improves vacuum performance in applica-

tions with condensable vapors.

Assemble the separator at the inlet using the

centring ring and the clamping ring.