• Ensure that the

coolant outlet tubing is always unob-

structed and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute. Outlet

flow must always be unhindered.

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or squeezed coolant hoses).

• Only install the optional coolant valve in the supply line

of the exhaust waste vapor condenser.

During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might gen-

erate overpressure at the outlet. Check pressure com-

patibility with system components (e.g., exhaust tubing

or exhaust valve) at the outlet. Ensure that the pump

outlet is neither blocked nor restricted.

+

Maximum ambient temperature: 104 °F (40 °C)

Check the maximum temperatures, if installing the

pump in a cabinet or a housing. Make sure ventilation

is adequate, especially if the ambient temperature is

elevated.

• If the pump is installed at an altitude of more than 6500

ft (2000 m) above mean sea level, check compatibility

with applicable safety requirements, and adopt suitable

measures. There is a risk of the motor overheating due

to insufficient cooling.

• Check compatibility with the

maximally permitted

pressure at outlet and the maximum pressure differ-

ence between inlet and outlet ports.