Technical data, Gas inlet temperatures, See „technical data“, pg – VACUUBRAND PC 8 m. RC 6 User Manual

Page 24: Protection, „technical data“, pg. 24)

Advertising
background image

page 24 of 49

See manual of • Diaphragm pump MD 12C or MV 10C or

• Chemistry-HYBRID-Pump RC 6

Type

PC 8 /

MD 12C

PC 8 /

MV 10C

PC 8 /

RC 6

Inlet

small flange KF 25

small flange

KF 16

Outlet

hose nozzle for tubing I.D. 3/8”

(hose nozzle DN 10 mm)

Coolant connection

hose nozzle for tubing I.D. 1/4” - 5/16”

(hose nozzle DN 6-8 mm)

Permissible range of coolant

temperature (waste vapor

condenser)

°F

(°C)

-

5 to 68

(-15 to +20)

Maximum permissible pres-

sure of coolant at waste vapor

condenser

psi

(bar)

87 (absolute)

(6 (absolute))

A-weighted emission sound

pressure level*

(uncertainty K

pA

: 3 dB(A))

dB(A)

52

52

50

Dimensions L x W x H

in

(mm)

20.0 x 14.4 x 16.9

(507 x 367 x 429)

20.0 x 14.8

x 16.9

(507 x 377 x

429)

Weight approx.

lbs. (kg) 69.4 (31.5) 69.4 (31.5) 68.8 (31.2)

Technical data

Gas inlet temperatures

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

* Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and ulti-

mate vacuum with exhaust tube at outlet.

Advertising