Permissible range of coolant temperature at the exhaust

waste vapor condenser: 5 °F to 68 °F (-15°C to +20°C)

Keep the coolant temperature low enough.

Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

Deposits inside the exhaust vapor condenser will reduce

its efficiency.

In case, clean the exhaust vapor condenser: Slacken the

pipe clamps, disconnect the hoses and remove the ex-

haust waste vapor condenser. Flush the exhaust waste

vapor condenser with a suitable cleansing agent.

Draining the separators:

Collecting flask at outlet:

Remove joint clip, remove collecting flask and drain con-

densate.

Separator at inlet (accessory; PC 8 with MD 12C / MV

10C):

Admit air or inert gas (via the pump inlet) to restore atmo-

spheric pressure in the separator. Open drain screw and

drain condensate.

Reattach drained collecting flask.

+

Important: Comply with regulations when disposing

of solvents/condensates. Recycle if possible; purify if

contaminated.