Keep the electrical power cord away from heated sur-

faces.

• Provide a firm, level platform for the equipment. En-

sure a stable position of the exhaust waste vapor con-

denser Peltronic™. Check that all fittings are secure.

• Comply with

maximum permissible pressures at in-

let and outlet and with

maximum permissible pres-

sure differences between inlet and outlet. See section

”Technical data”, pg. 22. Do not operate the exhaust

waste vapor condenser Peltronic™ with overpressure

at the inlet.

• Check the power source and the condenser’s rating

plate to be sure that the power source and the equip-

ment match in voltage, phase, and frequency.

• Make sure the fan’s air supply is adequate, especially

if the exhaust waste vapor condenser Peltronic™ is in-

stalled in an enclosure. Do not block ventilation slots.

• Connect hoses gas tight at inlet and outlet of the ex-

haust waste vapor condenser Peltronic™.

Note: Flexible elements will shrink when evacuated.

Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

(e.g., housing, walls, etc.), or else install an external auto-

matic ventilation system.

Check ventilation slots regularly for dust/dirt. Clean venti-

lation slots if necessary to avoid a reduction of ventilation.

Allow the equipment to equilibrate to ambient temperature

if you bring it from cold environment into a room prior to

operation. Notice if there is water condensation on cold

surfaces.