Troubleshooting – VACUUBRAND Exhaust waste vapour condenser Peltronic User Manual

Page 34

Advertising
background image

page 34 of 41

Troubleshooting

Fault

Possible cause

Remedy

❑ LED at the con-

denser is off, no fan

noise.

➨ Exhaust waste vapor

condenser Peltronic™

not switched on or con-

troller switched off?

✔ Switch on condenser and/

or controller.

➨ Electrical power cord

not plugged in, electri-

cal supply failure?

✔ Plug in power cord.

Check fuse.

➨ Exhaust waste vapor

condenser Peltronic™

defective?

✔ Contact local distributor.

❑ LED is flashing

orange, no change

to continuous green

light.

➨ Very high condensation

power?

✔ Check process parame-

ters (e.g., reduce pump-

ing speed of the vacuum

pump, increase cooling

at the vacuum side of the

system).

➨ Ambient temperature

too high?

✔ Comply with maximum

permissible ambient tem-

perature.

❑ Continuous red

light at the con-

denser.

➨ Ambient temperature

much too high?

✔ Comply with maximum

permissible ambient tem-

perature.

➨ Very high condensa-

tion power and ambient

temperature too high?

✔ Check process parame-

ters (e.g., reduce pump-

ing speed of the vacuum

pump, increase cooling

at the vacuum side of

the system). Comply with

maximum permissible

ambient temperature.

➨ Exhaust waste vapor

condenser Peltronic™

defective?

✔ Contact local distributor.