spheres.

Do not operate the exhaust waste vapor

condenser Peltronic™ in potentially explosive at-

mospheres.

Equipment without the

`

mark on the rating plate

is

not approved for the condensation of potentially

explosive atmospheres. Do not condense poten-

tially explosive atmospheres with this equipment.

Equipment bearing the

`

mark on its rating plate

is

approved for the condensation of potentially ex-

plosive atmospheres according to its classification

II 3G IIC T3 X according to ATEX, but it is not ap-

proved for operation in potentially explosive at-

mospheres (see section

`

Important information:

Equipment marking (ATEX)”, pg. 20)

.

Do not condense substances which may form depos-

its inside the exhaust waste vapor condenser Peltron-

ic™. The exhaust waste vapor condenser Peltronic™

is not suitable for condensing substances which may

form deposits inside the condenser. Deposits and con-

densate in the condenser may lead to increased tem-

peratures even to the point of exceeding the maximum

permitted temperatures.

• Check the inlet and outlet of the exhaust waste vapor

condenser Peltronic™, if there is a danger of forming

deposits inside the condenser. Inspect the exhaust

waste vapor condenser Peltronic™ regularly and clean

if necessary. See section ”Shutdown”, pg. 32).

Consider interactions and chemical reactions of

the condensing media. Ensure that the materials of

the condenser’s wetted parts are compatible with the

processed substances, see section ”Technical data”,

pg. 22.

When changing the substances processed, we rec-

ommend purging the exhaust waste vapor condenser

Peltronic™ with air or inert gas prior to changing the

Advertising