VACUUBRAND Exhaust waste vapour condenser Peltronic User Manual
Page 28
page 28 of 41
Hose cutter ....................................... 637257
Attention: The blade of the hose cutter is
very sharp. Risk of injury!
Connecting the outlet of the pump to the exhaust waste vapor con-
denser Peltronic™:
+
Establish connection with the formed PTFE hose included in delivery.
➨
Slip union nuts on hose. Fasten the PTFE hose with union nuts at the
tube fittings. If necessary widen the hose ends a bit by heating (538°F
(280°C) at maximum).
+
Depending on the pumping unit the formed
PTFE hose has to be cut accordingly.
Attention: Do not crush the PTFE hose,
in case use hose cutter (order number:
637257).
+
Pumping units comprising a diaphragm
pump with four pump heads and PC 2002:
Assemble formed hose as delivered (1).
+
Pumping units comprising diaphragm pumps
with two pump heads (except PC 2002): Cut
angled hose 90 mm x 60 mm (2).
+
MD 1C AK + EK: Cut straight piece of hose.
Length of 91 mm.
+
PC 2001: Cut straight piece of hose. Length:
85 mm. Mechanical fastening only possible
with special holding plate. Operation is also
possible with free-standing condenser.
(1)
(2)