Shutdown & storage, Notice – VACUUBRAND Exhaust waste vapour condenser Peltronic User Manual

Page 32

Advertising
background image

page 32 of 41

Shutdown & storage

Short-term:

Remove catchpot and drain condensate.

Important: Comply with regulations when disposing of

solvents/condensates. Recycle if possible; purify if con-

taminated.

Has the exhaust waste vapor condenser Peltronic™ been

exposed to media which may damage the condenser’s

materials or form

deposits?

Check the exhaust waste vapor condenser Peltronic™

(check inlet and outlet) and clean, if necessary.

Clean a contaminated exhaust waste vapor condenser

Peltronic™ as follows:

- Fill the condenser with a solvent (e.g., benzene) and

allow sufficient cleaning time. Observe all regulations

concerning usage and disposal of solvents!

- Drain the solvent and dispose of in accordance with

regulations. Repeat cleaning if necessary.

- Rinse the condenser several times with alcohol in or-

der to remove all solvent residues.

- Allow the emission condenser to dry.

Long-term:

- Take measures as described above regarding short-

term shutdown.

- Separate the exhaust waste vapor condenser

Peltronic™ from the application.

- Close inlet and outlet ports (e.g., with transport caps).

- Store the exhaust waste vapor condenser Peltronic™

under dry conditions.

NOTICE

Advertising