Exhaust waste vapor condenser Peltronic™ with level sensor
In case of high ambient humidity condensation may possibly occur on
the cold surfaces of the exhaust waste vapor condenser Peltronic™. If
condensation water is dripping onto the level sensor, an accidental level
sensor alarm might be triggered. When installing a level sensor at the
exhaust waste vapor condenser Peltronic™, pay attention to install it in a
way that no potential condensation water can drip onto the level sensor.
Avoid humidity on the sensor surface.
Valves and pressure transducers connected to the exhaust waste vapor
condenser Peltronic™ are handled just as well as components connect-
ed directly to the controller.
For optimal electromagnetic compatibility assemble a
ferrite at all VACUU•BUS connection cables (cat. no.:
612571).
NOTICE