Exhaust waste vapor condenser Peltronic™ with level sensor

In case of high ambient humidity condensation may possibly occur on

the cold surfaces of the exhaust waste vapor condenser Peltronic™. If

condensation water is dripping onto the level sensor, an accidental level

sensor alarm might be triggered. When installing a level sensor at the

exhaust waste vapor condenser Peltronic™, pay attention to install it in a

way that no potential condensation water can drip onto the level sensor.

Avoid humidity on the sensor surface.

Valves and pressure transducers connected to the exhaust waste vapor

condenser Peltronic™ are handled just as well as components connect-

ed directly to the controller.

For optimal electromagnetic compatibility assemble a

ferrite at all VACUU•BUS connection cables (cat. no.:

612571).