Never block the gas outlet. The exhaust hose has al-

ways to be unobstructed and without back pressure to

enable an unhindered discharge of gases.

+

Adopt suitable measures to prevent the release of dan-

gerous or polluting fluids. To prevent any emission of

such substances from the condenser outlet, install an

appropriate collecting and disposal system.

The electronic exhaust waste vapor condenser Peltron-

ic™ features a LED indicating the condenser’s operation

status (see also section ”Troubleshooting”).

Continuous green light: Condenser at operating tem-

perature (cold side < +54°F (+12°C)).

Flashing orange light: Condenser temperature elevated

(cold side > +54°F (+12°C)).

The condenser reaches its optimum cooling temperature

even without load only about one minute after switching on.

(LED: change from flashing orange to continuous green).

Continuous red light: Excess temperature at the con-

denser’s hot side. The cooling of the exhaust waste

vapor condenser Peltronic™ is switched off. The fan

remains running at maximum speed.

As soon as the condenser’s hot side has cooled down

to a temperature below 122°F (50°C), the exhaust

waste vapor condenser Peltronic™ restarts automat-

ically. The LED will be continuously cycling red and

orange respectively red and green (depending on the

temperature of the cold side). The flashing red light

indicates that a fault had occurred.

The fault indicator ”red light” can only be reset by

switching the condenser off/on. Determine and elimi-

nate the cause of failure (e.g., gas inlet temperature

too high, gas load too high, ambient temperature el-

evated).