processed media. Purging the exhaust waste vapor

condenser Peltronic™ will clean it from residues and it

will reduce the possibility of reactions of the processed

substances with each other and with the condenser’s

materials.

Safety during operation

Adopt suitable measures to prevent the release of dan-

gerous, toxic, explosive, corrosive, noxious or polluting

fluids, vapors and gases. To prevent any emission of

such substances from the condenser outlet, install an

appropriate collecting and disposal system and take

protective action for exhaust waste vapor condenser

Peltronic™ and environment.

You must take suitable precautions to prevent any for-

mation of explosive mixtures in the condenser or at its

outlet. In case, hot surfaces or static electricity may

ignite these mixtures.

Drain appropriately or otherwise remove any potential-

ly explosive mixtures at the outlet of the condenser, or

dilute them to non-explosive concentrations.

Never operate this equipment if it has a damaged cord

or plug.

+

If the condenser is not working properly, has been

dropped or has fallen into water, contact your service

provider.

+

Make sure that the exhaust pipeline cannot become

blocked.

+

Comply with applicable regulations when disposing of

chemicals. Take into consideration that chemicals may

be contaminated. Take adequate precautions to pro-

tect people from the effects of dangerous substances

(chemicals, thermal decomposition products of fluoro-