Operating conditions, Notice – VACUUBRAND MZ 2S NT User Manual

Page 16

Advertising
background image

page 16 of 91

To the best of our knowledge the equipment is in com-

pliance with the requirements of the applicable EC-direc-

tives and harmonized standards (see ”Declaration of Con-

formity”) with regard to design, type and model. Directive

EN 61010-1 gives in detail the conditions under which the

equipment can be operated safely (see also IP degree of

protection, “Technical data”, pg. 25).

Operating conditions

These pumps are not approved for operation in po-

tentially explosive atmospheres.

Do not operate the

pumps in potentially explosive atmospheres.

Pumps without the

`

mark on the rating plate are

not approved for the pumping of potentially explo-

sive atmospheres. Do not pump potentially explo-

sive atmospheres with those pumps.

Pumps bearing the

`

mark on their rating plates

are

approved for the pumping of potentially explo-

sive atmospheres according to their classification

II 3G IIC T3 X according to ATEX, but they are not

approved for operation in potentially explosive at-

mospheres (see section “

`

Important information:

Equipment marking (ATEX)”, pg. 23)

.

The pumps are

not suitable to pump any of the sub-

stances listed below.

Do not pump:

-

unstable substances

- substances which react explosively under

impact

(mechanical stress) without air

- substances which react explosively when being ex-

posed to

elevated temperatures without air,

- substances subject to

auto-ignition,

- substances which are

inflammable without air

-

explosive substances.

The pumps are

not approved for operation below

ground.

Do not operate the pump below ground.