VACUUBRAND MZ 2S NT User Manual

Page 22

Advertising
background image

page 22 of 91

mechanically stable. Check that the maximum gener-

ated overpressure is compatible with the mechanical

stability of the pressure vessel. Do not generate over-

pressure in vessels other than in those designed for

that purpose.

Risk of bursting!

Comply with

maximum permissible pressure at

the outlet (

58 psi (4 bar) absolute). The pump has

an overpressure safety relief device at the pressure

adjustment device at the outlet (opening pressure: 58

psi (4 bar) absolute). Install an additional overpressure

safety relief device in the pressure system, if neces-

sary.

Note: Flexible elements will expand when pressurized.

+

Especially when using the pump as vacuum pump pro-

vide a free and pressureless exhaust outlet to avoid

damage to pump valves and risk of bursting.

Advertising
This manual is related to the following products: