Notes prior to the use of the manometers (ME 4R NT)

The manometers are filled with glycerin. The overpressure manometer

indicates the overpressure relative to the atmospheric pressure at the

place of operation. The vacuum meter indicates the underpressure rela-

tive to the atmospheric pressure at the place of operation but with abso-

lute scaling.

+

Install the pump in the room of operation.

+

Vent the reference chambers prior to use. Else, not venting the ma-

nometers may lead to a systematic measuring error.

There are different types of manometers. To vent the reference chamber

check which types of manometers are installed at the pump and proceed

according to the given instructions:

A: Manometer with bore at the rear side

Pierce the hole at the rear side of the mano-

meter. Some outflow of liquid is normal.
+

Do not tilt pump during transport.

B: Manometer with pin

Lift the pin to vent the manometer. Do not re-

move the pin completely! Close after venting.
+

Repeat if necessary until the dial shows

ATM with the measuring connection at at-

mospheric pressure.

pin

bore

C: Manometer with pressure compensation

valve

Vent the manometer using the pressure com-

pensation valve. Close the valve after ventilation.
+

Repeat if necessary until dial shows zero

with measuring connection at atmospheric

pressure.

D: Manometer without bore, pin or pressure

compensation valve

Do

not vent the reference chamber of this ma-

nometer!

pressure

compensation

valve