Notice – VACUUBRAND MZ 2S NT User Manual

Page 58

Advertising
background image

page 58 of 91

Service kit for ME 2 NT................................................................................696877

Service kit for ME 4 NT / MZ 2 NT ..............................................................696860

Service kit for MD 4 NT / MV 2 NT .............................................................696861

Service kit for ME 8 NT .............................................................................. 696862

Service kit for MZ 2D NT ............................................................................696863

Service kit for ME 4S NT / MZ 2S NT ..........................................................696868

ME 8S NT / MD 4S NT ...............................................................................2x 696868

Service kit for ME 4R NT ............................................................................696859

Service kit for MD 4CRL NT ....................................................................... 696870

Diaphragm key (width 66 mm) ....................................................................636554

+

Allow sufficient cooling of the pump before

starting maintenance.

+

Vent the pump and isolate it from the

vacuum system before you start mainte-

nance.

Ensure that maintenance is done only by suitably

trained and supervised technicians.

The valves and diaphragms as well as the motor capac-

itors are wear parts. If the rated ultimate vacuum is no

longer achieved or in case of increased noise level, the

pump interior, the diaphragms and the valves must be

cleaned and the diaphragms and valves must be checked

for cracks or other damage.

All bearings are encapsulated and are filled with long-life

lubricant. Under normal operating conditions, the drive

system is maintenance free.

In demanding circumstances, it may be efficient to check

and clean the pump heads on a regular basis. In normal

use, the lifetime of the diaphragms and valves is typically

15,000 operating hours.

- Prevent internal condensation, transfer of liquids or

dust. The diaphragms and valves will be damaged if

liquid is pumped in significant amount.

- Carry out maintenance frequently if the pump is ex-

posed to corrosive media or in case of deposits.

- Regular maintenance will improve the lifetime of the

pump and also protect both users and the environment.

NOTICE

atm

°C / °F

Advertising
This manual is related to the following products: