Connecting the outlet

Outlet via silencer respectively via hose noz-

zle for tubing I.D. 3/8” (10 mm) (MD 4S NT) or

small flange KF DN 16 (MD 4CRL NT).

Attention: Use the silencer only in case of low

gas throughput. Check the silencer regularly

for permeability! In case, install a hose nozzle

with sealing ring instead (see “Accessories”,

pg. 51).).

Connecting an exhaust line at the outlet of a

pump with silencer at the outlet:

Unscrew the silencer and install a hose nozzle

DN 10 mm with sealing ring (thread G1/4”).

NOTICE

outlet

If dangerous or polluting fluids could be released at

the outlet, install an appropriate system to catch and

dispose of those fluids.

+

Connect a gas-tight exhaust line at the pump

outlet if necessary (in case, replace silencer

with hose nozzle, see above). Always vent

exhaust gases appropriately (e.g., into a

fume hood).

+

MD 4CRL NT: Connect a gas-tight exhaust

line at the pump outlet (small flange KF DN

16).

+

Never block the gas outlet. The exhaust

line must always be free of obstructions (no

back pressure) to ensure an unimpeded

discharge of gas. The cross-section of the

outlet tubing must be at least the size of the

pump’s exhaust connection.