Notice – VACUUBRAND MZ 2S NT User Manual

Page 47

Advertising
background image

page 47 of 91

• If the pump is installed at an altitude of more

than 6500 ft (2000 m) above mean sea lev-

el, check compatibility with applicable safety

requirements, and adopt suitable measures.

There is a risk of the motor overheating due

to insufficient cooling.

max. 6500 ft

(2000 m)

• Check compatibility with the

maximally

permitted pressure at inlet and at outlet

and the

maximum pressure difference

between inlet and outlet ports.

max. 16.0 psi

(1.1 bar) absolute

ME 4R NT:

Note: There are no stops at the end of the

dispensing screws’ threads! Do not com-

pletely unscrew the dispensing screws!

Do not start the pump if the

pressure difference between

inlet and outlet ports exceeds max. 16.0 psi (1.1 bar).

Attempts to start the pump at higher pressure difference

may cause stalling and damage of the motor.

NOTICE

Prevent internal condensation, transfer of liq-

uids or dust. The diaphragms and valves will

be damaged, if liquids are pumped in signifi-

cant amounts.

Check the pump regularly for external soiling

and deposits. Clean the pump if necessary

to avoid an increase of the pump’s operating

temperature.

NOTICE

IN

condensate / dust

Advertising