During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might

generate overpressure at the outlet. Check pressure

compatibility with system components (e.g., exhaust

tubing or exhaust valve) at the outlet. Ensure that the

pump outlet is neither blocked nor restricted.

max.

104°F

(40°C)

+

Maximum ambient temperature: 104 °F

(40 °C)

Check the maximum temperatures, if install-

ing the pump in a cabinet or a housing.

Make sure ventilation is adequate, especial-

ly if the ambient temperature is elevated.

+

Operation with silencer at the outlet: Oper-

ating the pump at a high inlet pressure or

pumping dusty gases for a long time may

cause clogging of the silencer. Check the

silencer regularly and replace if necessary,

or install a hose nozzle with sealing ring in-

stead (see “Accessories”, pg. 51).

ME 4R NT:

+

In case of continuous operation with inlet

pressures above atmospheric pressure: the

inlet pressure (p

in

) must not exceed the out-

let pressure (p

ex

).

Maximum permissible

outlet pressure: 58 psi (4 bar) absolute.

p

ex

> p

in

max. 43.5 psi (3 bar)

(relative)