The pigtail fuses are

integrated into wires

((1), black and blue)

inside the terminal

box. To replace the

fuses it is neces-

sary to replace both

wires completely

(fixed with flat pin

bushings (2)).

Replacing the fuse

Switch off the pump.

Disconnect the electrical power cord before open-

ing the terminal box. After disconnecting from power,

wait five seconds to allow the capacitors to discharge.

+

The replacing of the fuse has to be carried out by a

trained electrician.

+

After replacing the fuse, the pump must be checked for

electric safety (see below)! Identify and eliminate the

cause of failure before switching on the pump again.